54 II B Data Logging Thermometer with Dual Input | Fluke
Dansk

54 II B Data Logging Thermometer with Dual Input

  • Fluke 54 II B Dual Input Digital Thermometer with Data Logging

Nøglefunktioner

  • Large backlit dual display shows any combination of T1, T2 (52 and 54 only), T1-T2 (52 and 54 only), plus MIN, MAX, or AVG
  • Recall function allows logged data to be easily reviewed on the meter display
  • Infrared USB communication port allows data to be exported to optional FlukeView Forms® Temperature PC software for further analysis and graphing
  • Data Logging up to 500 points of data with user adjustable recording interval
  • Relative time clock on MIN, MAX, and AVG provides a time reference for major events
  • Electronic Offset function allows compensation of thermocouple errors to maximize overall accuracy
  • Measures J, K, T, E, R, S, and N-type thermocouples
  • Readout in °C, °F, or Kelvin (K)
  • Splash and dust resistant case protected by impact absorbing holster
  • User-friendly front panel is easy to set up and operate
  • Sleep mode increases battery life; typical 1000-hour battery life
  • Battery door allows easy battery replacement without breaking the calibration seal
  • Optional ToolPak accessory allows the thermometer to hang from any metal object (with the rare earth magnet) or secure around a pipe (with hook-and-loop straps) for hands-free operation
  • 3 year warranty

Produktoversigt: 54 II B Data Logging Thermometer with Dual Input

Laboratory accuracy. Wherever you go

Fluke 54 II B dual input digital thermometer can log up to 500 points of data to internal memory. Fast response and laboratory accuracy (0.05% + 0.3 °C). Take contact temperature for checking motors, insulation, breakers, pipes, corroded connections, liquids, and wires with industrial standard J, K, T, E, N, R, and S type thermocouple (temperature sensors).

Specifikationer: 54 II B Data Logging Thermometer with Dual Input

General Specifications
Temperature accuracyAbove -100°CJ, K, T, E, and N-type: ±[0.05% + 0.3°C]1
R and S-type: ±[0.05% + 0.4°C]1
Below -100°CJ, K, E, and N-types: ±[0.20% + 0.3°C]1
T-type: ±[0.50% + 0.3°C]
TemperatureJ-type-210°C to 1200°C
K-type-200°C to 1372°C
T-type-250°C to 400°C
E-type-150°C to 1000°C
N-type-200°C to 1300°C1
R and S-type0°C to 1767°C1
Temperature scaleITS-90
Applicable standardsNIST-175
Display resolution0.1°C, 0.1 K < 1000
1°C, 1 K ≥ 1000
1. Only the Fluke Models 53 II B and 54 II B thermometers are capable of measuring N, R, and S-type thermocouples.
Environmental Specifications
Operating temperature-10°C to 50°C
Storage temperature-40°C to 60°C
Humidity (without condensation)0% to 90%; 0°C to 35°C
0% to 70%; 0°C to 50°C
Safety Specifications
Overvoltage categoryCSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2
Agency approvalsCE, CSA, TÜV (pending)
Mechanical and General Specifications
Size (L x W x D)173 x 86 x 38 mm
Weight400 g
Batteries3 AA batteries; typical 1000-hour life

Modeller: 54 II B Data Logging Thermometer with Dual Input

Fluke 54 II
Fluke 54 II Dual Input Digital Thermometer
Køb

Includes:

  • Impact absorbing holster
  • Two 80PK-1 bead probe thermocouples