Fluke 37X Clamp Meter Recall Safety Notice | Fluke
English

Fluke 37X Clamp Meter Recall

Dear Fluke Customer:

Fluke is voluntarily recalling certain 373, 374, 375 and 376 Digital Clamp Meters that were manufactured between September 1, 2010 and October 31, 2012. The models and serial numbers affected are listed in the table below.

Model FamilyManufacture DatesStarting Serial No.Ending Serial No.
Fluke-376Sep 2010 – Feb 20111427000115909999
Fluke-375Sep 2010 – Mar 20111427000116379999
Fluke-374Oct 2010 – Feb 20111427000116079999
Fluke-373Oct 2010 – Oct 20121427000121950000

To determine if your clamp meter is affected by this recall notice, enter your model and serial number in the box below and click "Check." Any serial number preceded or followed by the letter "R" is not affected by this recall notice. The image below shows where to find the model and serial number of the affected clamp meters.

* If serial number is not available, your product is part of the recall. Follow the recall instructions below.

If you own one of these clamp meters, please stop using it and send it back to Fluke for repair. If you are not the primary user of the Fluke 37X Series Clamp Meter, please pass this notice along to the appropriate people within your organization.

(Please only send the clamp meter – any test leads/boxes/manuals that are sent with the unit will NOT be returned.)

Description of the problem

The printed circuit assembly may not be properly fastened to the test lead input jack. This may result in inaccurate voltage readings, including a low or no voltage reading on a circuit energized with a hazardous voltage, presenting a shock, electrocution or thermal burn hazard.

Actions

Immediately stop using your affected 37X series clamp meter(s) even if you have not experienced problems.

Contact Fluke for information on how to have your units repaired.

If you live in the United States

  • 1) Call (888) 98-FLUKE for information on how to return your unit for a free replacement.

    If you live in Canada call 1-800-36FLUKE (1-800-363-5853) for instructions on this recall.

    OR
     
  • 2) Call your local service number for information and to request a postage-paid mailing envelope be sent to you. (Please only send the clamp meter – any test leads/boxes/manuals that are sent with the unit will NOT be returned.) Specific country phone numbers are listed below.

    US : 1-877-564-7180

If you live in the EU

  • 1)  to register your affected product online as part of the recall. Once your information is complete and you click “submit,” you will receive an email with a printable, pre-paid and pre-addressed shipping label, a tracking number and a phone number for follow-up. Place the label on your shipping box and send to the address on the label. You should receive your unit back in 5 working days (1 week) from when it is shipped to Fluke. (Please only send the clamp meter – any test leads/boxes/manuals that are sent with the unit will NOT be returned.)

    OR
     
  • 2) Call your local service number and request an email with a printable, pre-paid and pre-addressed shipping label. (Please only send the clamp meter – any test leads/boxes/manuals that are sent with the unit will NOT be returned.) Specific country phone numbers are listed below.

    Austria : +43-1-928-9500
    Belgium : +32 2 402 21 00
    Denmark : +45 70 23 58 53
    Finland : +358 800-111862
    France : +33 170800000
    Germany : +49-69-2222-0200
    Italy : +39 02 36002000
    Netherland : +31 40-26751000
    Spain: +34 9141 40100
    Sweden : +46 8 566 37 400
    UK : +44-20 7942 0700

For all other parts of the world

We are undertaking this voluntary recall because we place tremendous importance on our customer's safety and the reputation of our products. Please accept our apologies for any inconvenience caused by this action.

Chat with ourFluke assistant
Clear Chat