主要功能
- Unlimited applications – See data sheet for powerful applications and details
- Built-in digital camera – Every Fluke Visual IR Thermometer has a built-in digital camera
- Exclusive thermal heat map blending overlay – With one button, blend digital images and thermal heat maps to instantly show the exact location of the issue
- No training required - Fully automatic with built-in intelligence
- True pocket size design – Less than half the size of an average infrared camera
- SmartView® reporting software – Store up to 10,000 images per Gb on included SD card and make professional reports on included professional software
- Automatic hot and cold spot trackers
- Breakthrough affordability – Fraction of the cost of infrared cameras that require costly training and software
- Powerful resolution with PyroBlend® Plus optics:
- Best-in-class field of view: 28 ° x 28 ° (40% wider for tight spaces)
- 4 times sharper images
- Alarm and time lapse features:
- Hi/Lo temperature alarm – Find hot spots in seconds with the alarm feature
- Time lapse image capture – Automatically monitor equipment over time
- Automated monitoring alarm – Automatically capture images after an alarm has been triggered
此產品已停產或重新分類
產品概述: VT04 Visual IR Thermometer 維護組合套件
VT04 與 Fluke 1507 絕緣電阻測試器是疑難排解與預防性維護應用的絕佳組合。
規格: VT04 Visual IR Thermometer 維護組合套件
Fluke VT04 Visual IR Thermometer
主要功能 | |
內建數位攝影機 | 有 |
熱像圖覆蓋功能 | 有 (5 種混合模式) |
影像光學系統 | PyroBlend® Plus 光學技術 *影像銳利度提升為 VT02 的 4 倍 |
可視範圍 | 28° x 28° |
高溫-低溫警示 | 有 |
延時影像擷取 | 有 |
自動監控警示 | 有 |
電池類型 | 充電式鋰電池 |
訓練 | 無需訓練便可上手 |
人體工學設計 | 纖巧的口袋大小設計 |
冷、熱標記 | 有 |
一般功能 | |
電池壽命 | 八 (8) 小時 |
溫度測量範圍 | -10 °C 到 +250 °C (14 °F 到 482 °F) |
溫度測量精確度 | +/- 2 °C 或 +/- 2 % |
溫度測量 | 有,中心點測量 |
儲存媒體 (Micro SD 卡) | 每 GB 可儲存高達 10,000 幅影像 (隨附四 (4) GB 卡) |
紅外線光譜波段 | 6.5 μm 至 14 μm |
水平和跨度 | 自動 |
對焦機制 | 無須對焦,且內建智慧型功能,不必接受任何訓練即可上手 |
對焦選項-獨家的使用者可選擇 NEAR/FAR (近端/遠端) 選項 | 近端 < 23 cm (9 in);遠端 > 23 cm (9 in) |
尺寸 | 21 cm x 7.5 cm x 5.5 cm (8.3 in x 3 in x 2.2 in);< 350 g. (12.5 oz.) |
檔案格式 | 以 .is2 格式儲存至 SD 記憶卡。使用者可在 SmartView 建立專業報告或匯出影像 (BMP、DIB、GIF、SPE、FIF、SPEG、JPG、PNG、TIF 和 TIFF) |
安全及法規遵循 | CFR47:2009 Class A. Part 15 subpart B;CE:EN 61326:2006;IEC/EN 61010-1:2010 |
保固 | 兩 (2) 年 |
Fluke 1507/1503 絕緣電阻測試器
交流/直流電壓測量 | ||
準確度 | 範圍 | 600.0 V |
解析度 | 0.1 V | |
50 Hz 到 400 Hz ± (% 的 Rdg + 位數) | ± (2 % ± 3) | |
輸入阻抗 | 3 MΩ (標稱值),< 100 pF | |
共模抑制比 (1 kΩ 非平衡) | 直流電為 50 或 60 Hz 時,大於 60 dB | |
過載保護 | 600 V 真均方根或直流 |
接地連接電阻測量 | ||
範圍/解析度 | 20.00 Ω | 0.01 Ω |
200.0 Ω | 0.1 Ω | |
2000 Ω | 1.0 Ω | |
20.00 kΩ | 0.01 kΩ | |
準確度 | ± (1.5 % ± 3) | |
過載保護 | 2 V 真均方根或直流 | |
開路測試電壓 | > 4.0 V,< 8 V | |
短路電流 | > 200.0 mA |
絕緣規格 | ||
測量範圍 | 1507 | 0.01 MΩ 到 10 GΩ |
1503 | 0.01 MΩ 到 2000 GΩ | |
測試電壓 | 1507 | 50 V、100 V、250 V、500 V、1000 V |
1503 | 500 V,1000V | |
測試電壓準確度 | + 20 %,- 0 % | |
短路電流 | 1 mA 標稱值 | |
自動 放電 | 電容 (C) = 1 µF 或更小時,放電時間小於 0.5 秒 | |
帶電電路指示器 | 在測試初始化前,如果端子電壓大於30 V,則禁止測試 | |
最大電容負載 | 1 µF 以下負載均可操作 | |
測量準確度 | 50 V | ± (3 % ± 5) |
100 V | ± (3 % ± 5) | |
250 V | ± (1.5 % ± 5) | |
500 V | ± (1.5 % ± 5) | |
1000 V | 2000 MΩ 以內為 ±(1.5 % + 5),超過 2000 MΩ 為 ±(10 % + 3) |
一般規格 | ||
任何端子可承受的最大電壓 | 600 V 交流真均方根或直流 | |
存放溫度 | -40 °C 到 60 °C (-40 °F 到 140 °F) | |
操作溫度 | -20 °C 到 55 °C (-4 °F 到 131 °F) | |
溫度係數 | 溫度 < 18 °C 或 > 28 °C (< 64 °F 或 > 82 °F) 時,每 °C 為 0.05 x (指定精確度) | |
相對濕度 | 10 °C 到 30 °C (50 °F to 86 °F) 下為 0 % 到 95 % | |
30 °C 到 40 °C (86 °F to 104 °F) 下為 0 % 到 75 % | ||
40 °C 到 55 °C (104 °F to 131 °F) 下為 0 % 到 40 % | ||
振動 | 隨機,2 g,5-500 Hz (根據 MIL-PRF-28800F Class 2 儀器) | |
衝擊 | 可承受自 1 公尺高處落下,根據 IEC 61010-1 第二版 (1 公尺摔落測試,六面,橡木地板) | |
電磁相容性 | 在 3 V/M 的無線電頻率場中,準確度 = 指定準確度 (EN 61326-1:1997) | |
安全性 | 符合 ANSI/ISA 82.02.01 (61010-1) 2004、CAN/CSA-C22.2 NO. 61010-1-04,與針對測量類別 IV 600 V (CAT IV) 的 IEC/EN 61010-1 第二版 | |
認證 | 根據標準 CSA/CAN C22.2 No. 61010.1-04 的 CSA;根據標準 IEC/EN 61010-1 第二版的 TUV | |
電池 | 四個 AA電池 (NEDA 15A 或 IEC LR6) | |
電池壽命 | 絕緣測試使用 | 室溫條件下,新的鹼性電池可供測試器進行至少 1000 次絕緣測試。此為在 1000 V、1 MΩ 及 5 秒啟動和 25 秒關閉的工作週率下進行的標準測試。 |
電阻測量 | 室溫條件下,新的鹼性電池可供測試器進行至少 2500 次接地連接電阻測量。此為在1 MΩ 及 5 秒啟動和 25 秒關閉的工作週率下進行的標準測試。 | |
尺寸 (長x 寬x 高) | 高 5.0 cm x 寬 10.0 cm x 長 20.3 cm (高 1.97 in x 寬 3.94 in x 長 8.00 in) | |
重量 | 550 g (1.2 lb) | |
IP 評等: | IP40 | |
高度 | 操作 | 2000 m CAT IV 600 V,3000 m CAT III 600 V |
非操作中 (存放) | 12,000 m | |
超過範圍耐受能力 | 範圍的 110 % | |
隨附配件 | TL224 測試導線、TP74 測試探針、PN 1958654 (紅) 與 PN 1958646 (黑) 夾具、保護套與遠距探針 |
機型: VT04 Visual IR Thermometer 維護組合套件
Fluke VT04-Maint-Kit
Fluke VT04 Visual IR Thermometer Maintenance Combo Kit
隨附內容:
- VT04 Visual IR Thermometer
- 1507 絕緣電阻測試器