Fluke 754 Documenting Process Calibrator-HART | Fluke
English

Fluke 754 Documenting Process Calibrator-HART

Fluke 754 Documenting Process Calibrator-HART is rated 3.9 out of 5 by 15.
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 24.5ms
  • REVIEWS, PRODUCT
  • Fluke 754 Documenting Process Calibrator-HART
  • Fluke 754 Documenting Process Calibrator-HART
  • Fluke 754 Documenting Process Calibrator-HART
  • SS: 753 Documenting Process Calibrator

Key features

  • Complete pressure, temperature, and mA loop calibrator
  • Troubleshoot and calibrate HART smart digital transmitters
  • Create calibration procedures and automatically document results
  • Connect to calibration management software

Product overview: Fluke 754 Documenting Process Calibrator-HART

Fluke 754 Documenting Process Calibrator with HART communication does the work of many tools

Whether you're calibrating instruments, troubleshooting a problem, or running routine maintenance, the Fluke 754 with HART® communication can help you get the job done faster. It does so many different tasks, so quickly and so well, it's the only process calibrator you need to carry. This rugged, reliable integrated communicating calibrator is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.

The 754 is a power multifunction documenting calibrator that you can use to download procedures, lists, and instructions created with software; or upload data for printing, archiving, and analysis. The powerful built-in HART interface is capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.

In fact, the 754 does the work of several tools. It sources, simulates and measures pressure, temperature, and electrical signals with one rugged, hand-held device. For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations. Plus, the new improved graphical screen, Li-Ion battery for longer life, USB port, and accessories help you work smarter and faster.

To create a seamless/paperless calibration management system consider adding Fluke DPCTrack2 Calibration Management software for use with your Fluke 753 and 754 or even legacy Fluke 743 and 744 calibrators.

Other useful features:

  • Handles fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms
  • Measures/sources pressure using any of 29 Fluke 700Pxx Pressure Modules
  • Creates and runs automated as-found/as-left procedures to satisfy quality programs or regulations, and records and documents results
  • Holds up to a full week of downloaded procedures and calibration results
  • Uses many features like autostep, custom units, user-entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay and more
  • Provides easy-to-use multi-lingual interface
  • Features a bright white dual display for reading both sourced and measured parameters simultaneously
  • Includes rechargeable Li-Ion battery for 10 hour uninterrupted use
  • Comes with three-year warranty and DPC/Track Sample software
  • Offers compatibility with many asset management software packages

Related blog posts:

Calibrating a HART temperature transmitter

What's in the box:

  • Fluke 754 Documenting Process Calibrator-HART
  • BC7240 battery charger
  • Li-on BP7240 battery pack
  • NIST-traceable calibration report and data
  • (3 sets)TP220 test probes
  • (3 sets) "extended tooth" alligator clips
  • (2 sets) AC280 hook clips
  • C799 Soft Field Case
  • USB communication cable, Fluke 754HHC HART communications cable
  • Quick Reference Guide
  • 3-year warranty

Specifications: Fluke 754 Documenting Process Calibrator-HART

Measurement Accuracy
Voltage DCRange/ Resolution1 Year2 Years
100.000 mV0.02% + 0.005 mV0.03% + 0.005 mV
3.00000 V0.02% + 0.00005 V0.03% + 0.00005 V
30.0000 V0.02% + 0.0005 V0.03% + 0.0005 V
300.00 V0.05% + 0.05 V0.07% + 0.05 V
Voltage AC3.000 V (40 Hz to 500 Hz) / 0.001 V0.5% + 0.002 V1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V0.5% + 0.02 V1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V0.5% + 0.2 V1.0% + 0.2 V
Current DC30.000 mA0.01% + 5 uA0.015% + 7 uA
110.00 mA0.01% + 20 uA0.015% + 30 uA
Resistance10.000 Ω0.05% + 50 mΩ0.07% + 70 mΩ
100.00 Ω0.05% + 50 mΩ0.07% + 70 mΩ
1.0000 kΩ0.05% + 500 mΩ0.07% + 0.5 Ω
10.000 kΩ0.1% + 10 Ω0.15% + 15 Ω
Frequency1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz 0.05 kHz
Source Accuracy
  1 Year2 Years
Voltage DC100.000 mV0.01% + 0.005 mV0.015% + 0.005 mV
1.00000 V0.01% + 0.00005 V0.015% + 0.0005 V
15.0000 V0.01% + 0.0005 V0.015% + 0.0005 V
Current DC22.000 mA (source)0.01% + 0.003 mA0.02% + 0.003 mA
Current sink (simulate)0.02% + 0.007 mA0.04% + 0.007 mA
Resistance10.000 Ω0.01% + 10 mΩ0.015% + 15 mΩ
100.00 Ω0.01% + 20 mΩ0.015% + 30 mΩ
1.0000 kΩ0.02% + 0.2 Ω0.03% + 0.3 Ω
10.000 kΩ0.02% + 3 Ω0.03% + 5 Ω
Frequency0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Technical Data
Data log functionsMeasure functionsVoltage, current, resistance, frequency, temperature, pressure
Reading rate1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length8000 readings (7980 for 30 or 60 readings/minute)
Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
Rate4 steps/second
Trip detectContinuity or voltage (continuity detection not available when sourcing current)
Loop power functionVoltageSelectable, 26 V
Accuracy10%, 18 V minimum at 22 mA
Maximum current25 mA, short circuit protected
Maximum input voltage50 V DC
Step functionsSource functionsVoltage, current, resistance, frequency, temperature
Manual stepSelectable step, change with arrow buttons
AutostepFully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature-10°C to +50°C
Storage temperature-20°C to +60°C
Dust/water resistanceMeets IP52, IEC 529
Operating altitude3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvalsCAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight1.2 kg (2.7 lb)
BatteriesInternal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Battery life>8 hours typical
Battery replacementReplace without opening calibrator; no tools required
Side port connectionsPressure module connector
USB Connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Data storage capacity1 week of calibration procedures results
90 day specificationsThe standard specification interval for the 750 Seriesare 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
Temperature, Resistance Temperature Detectors
Degrees or % of reading - Type (α)Range °CMeasure °C1
 1 year2 year
100 Ω Pt (385)-200 to 100
100 to 800
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
200 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
500 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
1000 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3916)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3926)-200 to 100
100 to 630
0.08°C
0.02% + 0.06°C
0.16°C
0.04% + 0.12°C
10 Ω Cu (427)-100 to 2600.2°C0.4°C
120 Ω Ni (672)-80 to 2600.1°C0.2°C
Source currentSource °CAllowable current2
 1 year2 year 
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
500 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
250 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
150 μA0.06 C
0.017% + 0.05°C
0.12 C
0.034% + 0.10°C
0.1 mA to 1 mA
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
3 mA0.2°C0.4°C0.1 mA to 10 mA
1 mA0.04°C0.08°C0.1 mA to 10 mA
1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
Temperature, Thermocouples
TypeSource °C Measure °C Source °C
  1 year2 years1 year2 years
E-250 to -2001.32.00.60.9
-200 to -1000.50.80.30.4
-100 to 6000.30.40.30.4
600 to 10000.40.60.20.3
N-200 to -1001.01.50.60.9
-100 to 9000.50.80.50.8
900 to 13000.60.90.30.4
J-210 to -1000.60.90.30.4
-100 to 8000.30.40.20.3
800 to 12000.50.80.30.3
K-200 to -1000.71.00.40.6
-100 to 4000.30.40.30.4
400 to 12000.50.80.30.4
1200 to 13720.71.00.30.4
T-250 to -2001.72.50.91.4
-200 to 00.60.90.40.6
0 to 4000.30.40.30.4
B600 to 8001.32.01.01.5
800 to 10001.01.50.81.2
1000 to 18200.91.30.81.2
R-20 to 02.32.81.21.8
0 to 1001.52.21.11.7
100 to 17671.01.50.91.4
S-20 to 02.32.81.21.8
0 to 2001.52.11.11.7
200 to 14000.91.40.91.4
1400 to 17671.11.71.01.5
C0 to 8000.60.90.60.9
800 to 12000.81.20.71.0
1200 to 18001.11.60.91.4
1800 to 23162.03.01.32.0
L-200 to -1000.60.90.30.4
-100 to 8000.30.40.20.3
800 to 9000.50.80.20.3
U-200 to 00.60.90.40.6
0 to 6000.30.40.30.4
BP0 to 10001.01.50.40.6
1000 to 20001.62.40.60.9
2000 to 25002.03.00.81.2
XK-200 to 3000.20.30.20.5
300 to 8000.40.60.30.6

Models: Fluke 754 Documenting Process Calibrator-HART

Fluke 754
Fluke 754 Documenting Process Calibrator-HART

Includes:

  • BC7240 battery charger
  • Li-on BP7240 battery pack
  • DPCTrack 2™ Sample Software
  • Instruction manual
  • NIST-traceable calibration report and data
  • Three sets of TP220 test probes with three sets of "extended tooth" alligator clips
  • Two sets AC280 hook clips
  • C799 Soft Field Case
  • USB communication cable, Fluke 754HHC HART communications cable

Is it possible to use this communicator to setup up a HART device? We tried to use this for a device setup but were not able. All that could really be done is read the device outputs which is almost useless.

Asked by: Very Dissapointed
The Fluke 754 Documenting Process Calibrator-HART is a calibrator and only supports calibration related HART functions. The Fluke 154 HART Communicator is a HART communicator and can be used to fully commission a HART capable instrument.
Answered by: FlukeCorp
Date published: 2024-10-13
  • y_2024, m_11, d_3, h_7CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_US, sid_3739311570996, prod, sort_[SortEntry(order=LAST_APPROVED_ANSWER_SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_fluke
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 24.39ms
  • QUESTIONS, PRODUCT
Rated 3 out of 5 by from Limited HART Capability The Fluke 754 does not have full handshake HART communication abilities. Not only is the device list short with regard to what instruments it can communicate with, the 754 cannot access all menus or parameters made available by the given device. If you must stay within the Fluke product line and require a documenting unit for basic data generation, purchase the 753, and utilize your pc, tablet or handheld for communications.
Date published: 2024-03-09
Rated 5 out of 5 by from Versatile I bought the Fluke 754 to perform calibrations on furnace thermocouples. Since the meter can do so many other measurements, I am investigating performing additional calibrations including pressure meters.
Date published: 2016-12-12
Rated 5 out of 5 by from Great tool! I purchased this tool a little over two years ago now and still going strong getting beat around out in the oil fields. It's an indispensable part of my artillery for maintaining and repairing gas compressor automation systems and pretty much anything else I've run across
Date published: 2016-12-08
Rated 3 out of 5 by from Fluke 754 Today I experienced a little issue that made me had to go back to the shop and get a 475. While doing a functional check/calibration of a Rosemount 3051 at 25 inH2O and using an ADT761 to run the the control the 754 would not let me adjust the transmitter without a Fluke pressure module. So I went back the the shop and got the 475 and left the Fluke.
Date published: 2016-12-08
Rated 4 out of 5 by from Definitely an upgrade but still bulky Much better than my 744. easier to see and battery life is much longer. was hoping they would make these smaller. the hart cable is much better. not a fan of the connection they use for presure transmitters but glad they kept it the same so i dont have to get new modules!
Date published: 2016-12-07
Rated 5 out of 5 by from Great for educators!!! We bought four Fluke 754 process calibrators this year. The versatility of this meter has proven to be the best feature for educating future instrumentation technicians. From pressure and temperature switches to all types of transmitters in our measurements classes and labs, the Fluke 754 is an all around centerpiece of the lab. In our process instrumentation class and labs, we introduce the HART capability of our instruments and the Fluke 754 is again the most versatile calibrator. In our calibration class is where the Fluke 754 has its best performance as the double screen lets the students have total control of the calibration process.
Date published: 2016-12-07
Rated 5 out of 5 by from Perfect Do It All Meter! I have been using this meter for over a year. I am able to do anything that I need to do to troubleshoot and calibrate industrial instrumentation. This meter eliminates the need to carry multiple calibration tools into the field. The HART function is also very handy! Best all around meter I own!
Date published: 2016-12-07
Rated 4 out of 5 by from Great unit!! Include a mA clamp option!! Excellent Meter, however fluke should provide a plug in mA current probe option so loop measurements can be done done without breaking the circuit!! Saves me carrying around the 773 as well.
Date published: 2014-12-15
  • y_2024, m_11, d_4, h_7
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_15
  • loc_en_US, sid_3739311570996, prod, sort_[SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_fluke
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 24.01ms
  • REVIEWS, PRODUCT
Please enable cookies to view product reviews.