Fluke 85RF High Frequency Probe | Fluke
ภาษาไทย

ภาพรวมผลิตภัณฑ์: Fluke 85RF High Frequency Probe

The 85RF High Frequency Probe is designed to convert a DC voltmeter with a input impedance of 10M ohms into a high frequency (100 kHz to 500 Mhz) RF voltmeter. Conversion from AC to DC is accomplished on a one-to-one basis includes a range of 0.25 to 30 Vrms. The probe's DC output is calibrated to be equivalent to the rms value of a sine wave input.

ข้อมูลจำเพาะ: Fluke 85RF High Frequency Probe

Specifications
Voltage range0.25 to 30 Vrms
Maximum input voltage30 Vrms, 200 VDC
AC to DC ratio1:1
Ratio Accuracy (At 25 MHz and loaded with 10 M ohm) Above 0.5 V±0.5 dB
Below 0.5 V: ±1.0 dB
Add ±0.2 dB in RF fields of 1 to 3 V/m
Frequency response (relative to 25 MHz)100 KHz to 100 MHz ±0.5 dB
100 MHz to 200 MHz ±2.0 dB
200 MHz to 500 MHz ±3.0 dB

รุ่น: Fluke 85RF High Frequency Probe